18
Table3:Guidanceandmanufacturer’sdeclaraon
-electromagnecimmunity
TheSolisPainReliefSystemisintendedforuseintheelectromagnecenvironmentspecied
below.Thecustomerorusershouldassurethatitisusedinsuchanenvironment.
Immunity
test
IEC 60501
Testlevel
Compliance
level
Electromagnecenvironmentguidance
PortableandmobileRFCommunicaons
equipmentshouldbeusednoclosertoany
partofthedevice,includingcables,thanthe
recommendeddistancecalculatedfromthe
equaonapplicabletothefrequencyofthe
transmier’srecommendedseparaondistance
Radiated
RFIEC
61000-4-3
3V/m
80MHzto
2.5GHz
3V/m
80MHzto800MHz
d=1.2√P
800MHzto2.5GHz
d=2.3√P
Where“P”isthemaximumoutputpowerrang
ofthetransmierinWas(W)accordingto
thetransmiermanufacturerand“d”isthe
recommendedseparaondistanceinmeters
“m”.FieldstrengthfromxedRFtransmiersas
determinedbytheelectromagnecsitesurvey,a.
shouldbelessthanthecompliancelevelineach
frequencyrange.b.interferencemayoccurinthe
vicinityofequipmentmarkedwiththefollowing
symbol:
NOTE1:At80MHzto800MHz,thehigherfrequencyrangeapplies.
NOTE2:Theseguidelinesmaynotapplyinallsituaons.Electromagnecpropagaonisaectedbyabsorpon
andreeconfromstructures,objectsandpeople.
a.Fieldstrengthsfromxedtransmiers,suchasbasedstaonsforradio(cellularandcordless)andland
mobileradios,amateurradio,AMandFMradiobroadcast,andTVbroadcastcannotbepredictedtheorecally
withaccuracy.ToassesstheelectromagnecenvironmentduetoxedRFtransmiers,anelectromagnecsite
surveyshouldbeconsidered.IfthemeasuredeldstrengthinthelocaoninwhichtheSolisPainReliefSystem
isusedexceedstheapplicableRFcompliancelevelabove,theSolisPainReliefSystemshouldbeobservedto
verifynormaloperaon.Ifabnormalperformanceisobserved,addionalmeasuresmaybenecessary,suchas
reorienngorrelocangthedevice.
b.Overthefrequencyrangeof150KHzto80MHz,eldstrengthsshouldbeless
than3V/m.