Table2:Guidanceandmanufacturersdeclaraon
-electromagnecimmunity
TheSolisPainReliefSystemisintendedforuseintheelectromagnec
environmentspeciedbelow.Thecustomerortheuser
shouldassurethatitisusedinsuchenvironment
Immunity
test
IEC 60601
testlevel
Compliance
level
Electromagnec
Environment
guidance
Electrostac
Discharge
(ESD)IEC
61000-4-2
±6KV
Contact
±8KV
air
±6KV
Contact
±8KV
air
Floorsshouldbewood,
concrete,orceramicle.
Ifoorsarecoveredwith
synthecmaterial,the
relavehumidityshould
beatleast30%
PowerFrequency
(50/60Hz)
Magnec
Field
IEC61000-4-8
3A/m 3A/m
Powerfrequencymag-
neceldsshouldbeat
levelscharacteriscofa
typicallocaoninatypical
commercialorhospital
environment