Do you have a question about the Sony SH800 and is the answer not in the manual?
Type | Cell Sorter |
---|---|
Software | Sony SH800 Software |
Category | Laboratory Equipment |
Sample Input | Tubes or plates |
Laser Options | Up to 4 lasers |
Clean collection chamber and sort collection device. Use 70% EtOH on plates.
Ensure filled sheath tank, de-bubble filter for stable stream.
Activate pressure before proceeding with subsequent steps.
Turn on the SH800 instrument and observe fluid levels on display.
Disconnect airline, release pressure, and fill sheath tank carefully.
Empty waste tank at sink, add bleach to carboy before replacing.
Use standby mode for emptying waste or filling sheath to avoid recalibration.
Fill Di water tank, then de-bubble the filter using software.
Log into the Windows operating system using the FACS User account.
Launch the SH800 software after the instrument enters standby mode.
Log into the SH800 software with user credentials; use 'facs user' if needed.
Enable aerosol management system for BSL-2 samples.
Fill out the biohazard form detailing sample hazard and cleanup.
Scan and load a new chip; reuse old chip if <24h old.
Select blue laser for calibration, then run autocalibration with beads.
All samples, including compensation controls, must be filtered.
Select experiment, name, and configure fluorescent parameters.
Turn on 488nm laser; use violet laser sparingly to preserve chip life.
Use compensation wizard or manual method with single-color controls.
Adjust FSC/BSC gain and set detector thresholds for accurate cell acquisition.
Acquire sufficient events for setting gates and gain adjustments.
Create plots, double-click gates, and adjust settings for desired populations.
Select sort mode, set sort gates, place tubes, and start sorting.
Install splash guard, select 96-well plate method, and pause for gate setup.
Adjust plate drop positions, verify cell distribution, and re-adjust if needed.
Clear wells, set sort criteria, click 'Sort and Record Start'.
Click Worksheet Tools and 'Analyze Index Data' to review sort results.
Put the machine in standby mode via settings before maintenance tasks.
Export FCS files and save PDF of experiment layout and gates.
Perform bleach clean and DI water rinse cycles after use.
Log out, checking 'Keep sort calibration' to avoid recalibration.
Use aerosol management system, keep sort door closed, and manage suction.
Perform hardware/software shutdown, bleach clean, DI rinse, and vent sheath tank.
Guidance on picking dyes for optimal excitation with available lasers.
List of dyes compatible with SH800 sorter's lasers and their excitation.
Note on no RED laser and sub-optimal excitation for certain dyes with yellow-green laser.