5.Press [6] to select Standard Curve menu item. Press [Enter].The LCD will display as blow.
6.Press [1] to select Create Curve menu item. Press [Enter].The LCD will display as blow.
7.Press [1] to select No. of samples menu item. Press [Enter].The LCD will display as blow.
Input
Sample Num:
1No. of Samples
2Unit
3Input Conc
4Test Samp le
1Create Curve
2Load Curve
3Delete Curve
4Print Curve
5Lamp on/off
6Standard Curve
7Language
8Com. With PC
8. Use digital key to enter samples number, press [Enter].
9. Press [2] to select Unit menu item. Press [Enter].The LCD will display as blow.
10.
Use digital key to enter sample unit, press [Enter].
11. Press [3] to select Input Conc menu item. Press [Enter].The LCD will display as blow.
Input
Conc:
Number:1
1mg/L
2ug/L
3ng/kg
4ug/kg
12. Use digital key to enter sample concentration, press [Enter].
13. Press [4] to select Test Sample menu item. Press [Enter].The LCD will display as blow.
Number: 1
Conc: 1.00
Abs.:
14.Insert the standard solutions in the light beam. Press [Enter].
Users Manual SP-UV 200 22
Spectrum Instruments

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Spectrum SP-UV 200 and is the answer not in the manual?

Spectrum SP-UV 200 Specifications

General IconGeneral
BrandSpectrum
ModelSP-UV 200
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals