What to do if the audio alarm is sounding and the 'Low Pressure' light is flashing on my Squirrel Medical Medical Equipment?
- EEric ReynoldsAug 3, 2025
If the audio alarm is sounding and the 'Low Pressure' light is flashing, it could be due to a couple of reasons. First, check if the mains power is connected. If you want the Dormir Digital MRS to be on, reconnect the mains power. If you want it off, press the 'On/Off' button on the control panel until the unit switches off. Another cause could be a leak in the system. Check that the tube set is fully connected to the pump, the CPR valve is fully sealed, the 'O' rings on the tube set connector are present and in good condition, and that the mattress or tube set hasn't been punctured. The alarm will stop when the system pressure reaches a safe level.