Squirrel Dormir Digital MRS
Instructions for Use
IMPORTANT!
Read the instructions in this booklet thoroughly prior to use.
Keep these instructions for future reference.

Questions and Answers

  • F
    Fernando CampbellAug 2, 2025
    What to do if the audio alarm is sounding and the 'Low Pressure' light is flashing on my Squirrel Medical Medical Equipment?
    • E
      Eric ReynoldsAug 3, 2025
      If the audio alarm is sounding and the 'Low Pressure' light is flashing, it could be due to a couple of reasons. First, check if the mains power is connected. If you want the Dormir Digital MRS to be on, reconnect the mains power. If you want it off, press the 'On/Off' button on the control panel until the unit switches off. Another cause could be a leak in the system. Check that the tube set is fully connected to the pump, the CPR valve is fully sealed, the 'O' rings on the tube set connector are present and in good condition, and that the mattress or tube set hasn't been punctured. The alarm will stop when the system pressure reaches a safe level.
  • Y
    ygutierrezAug 7, 2025
    Why is the audio alarm sounding at regular intervals on my Squirrel Medical Medical Equipment?
    • L
      Linda JohnsonAug 7, 2025
      The audio alarm sounding at regular intervals could be due to a leak in the system or an incorrect pressure setting. Check the following: Is the CPR valve fully sealed? Are the 'O' rings on the tube set connector present and in good condition? Has the mattress or tube set been punctured? Also, ensure that the 'Comfort Level' is set appropriately for the mattress occupant.
  • D
    Daniel HernandezAug 12, 2025
    What to do if my Squirrel Medical Medical Equipment pump will not switch on?
    • D
      Diane SchultzAug 12, 2025
      If the pump will not switch on, check the following: Is the mains cable undamaged? Is the pump fully plugged into the mains socket and switched on? Have the fuses in the pump blown? Has the mains plug fuse blown?
  • R
    Robert CochranAug 18, 2025
    What to do if cells in mattress are not alternating on my Squirrel Medical Medical Equipment?
    • R
      Renee BaileyAug 18, 2025
      If the cells in the mattress are not alternating, it might be because 'Static' mode is enabled. Press the 'Static/Alternating Mode' button so that the 'Alternating' light illuminates.
  • K
    kim62Aug 22, 2025
    What to do if the 'Ready' light has not lit within 30 minutes of switching on my Squirrel Medical Medical Equipment?
    • F
      faithholmesAug 23, 2025
      If the 'Ready' light hasn't lit within 30 minutes of switching on, there may be a leak in the system. Check that the tube set is fully connected to the pump, the CPR valve is fully sealed, the 'O' rings on the tube set connector are both present and in good condition, and whether the mattress or tube set has been punctured.
  • C
    Corey FosterAug 28, 2025
    What does it mean if the 'Low Pressure' light is occasionally blinking (no audio alarm) on my Squirrel Medical Medical Equipment?
    • D
      Dawn TateAug 28, 2025
      If the 'Low Pressure' light is occasionally blinking without an audio alarm, it could indicate an incorrect pressure setting. Ensure that the 'Comfort Level' is set to an appropriate level for the mattress occupant.

Summary

2. Intended Use

2.1 Contra Indications

Details conditions where the Dormir Digital MRS should not be used, particularly spinal injuries.

2.2 General Warnings

Provides crucial safety warnings regarding modification, usage, patient transfer, and proximity to medical equipment.

4. Installation

5. Dormir Digital MRS Operation

5.1 Pre-Use Routine

Checklist of essential steps to perform before each use to ensure correct and safe operation of the MRS.

5.2 Control Panel Overview

Detailed explanation of the pump's control panel buttons, indicators, and their functions.

5.3 Using the Dormir Digital MRS

Instructions on how to turn on the pump, inflate the mattress, and when the MRS is ready for use.

5.4 Comfort Settings and Modes

Guidance on adjusting comfort levels, using Static Mode, and weight limits for the MRS.

5.5 Alarms and Indicators

Explanation of the Power Failure Alarm, its causes, and troubleshooting steps.

5.6 Low Pressure Alarm

Describes the audible alarm for low pressure, its causes, and checks for leaks or CPR valve issues.

5.7 CPR Valve

Explains the function and operation of the CPR valve for rapid mattress deflation in emergencies.

6. Storage and Transport

7. Cleaning and Decontamination

9. Troubleshooting

9.1 Pump Not Switching On

Diagnosing and resolving issues where the pump fails to power on, checking mains power and fuses.

9.2 No Ready Light

Troubleshooting steps for when the 'Ready' light does not illuminate, focusing on system leaks and connections.

9.3 Audio Alarm and Low Pressure Light

Addressing audio alarms and low pressure lights, checking power, leaks, and pressure settings.