Titan S:
Tube Assembly, Gasket, Retainer Ring, (2/3-Line) ..........................256454
Tube Assembly, Gasket (4-Line) .............................................................256455
Instruction Manual
Titan
®
Blis-Sonic Scaler SW/S
Replacement Assemblies and Spare Kits
WARNING: Never remove scaler housing with tip assembled to the scaler.
1. Unthread the front of the scaler housing to expose the front suspension carrier.
2. Remove the two old O-rings and install the replacement O-rings, StarDental Part #256718.
3. Re thread front of scaler housing to original position.
Titan S/SW Rear Seal & Swivel O-Rings
O-Rings & Rear Seal Replacement
Swivel O-Rings: Periodically replace the swivel
O-rings, StarDental
®
Part #259884.
Front Suspension O-Rings: After two years, front
suspension O-rings relax resulting in play between
shaft and handle.
Swivel O-Rings
Front Suspension O-Rings
Front of Scaler Housing
Seals wear with repeated use. Replace rear seals
annually or as needed. On either the
2/3-line swivel or Titan S 2/3-line Scaler replace
the 2/3-line rear seal with StarDental Part #203103.
On either the 4-line swivel or Titan S 4-line Scaler
replace seal with StarDental Part #202897.
Rear Seal on
the Swivel
Rear Seal on
the 4-line
Titan S
2/3-line for Titan S/SW
1. Remove 2/3-line rear seal with an
explorer.
2. Place new rear seal over drive air and
water tubes.
3. Carefully push seal into place using a
blunt instrument.
Rear Seal
Titan S 2/3-line
Titan SW
2/3-line
Rear Seal
Swivel
Cleaning Wire
CLEANING the Tips
Clean tips weekly by running cleaning wire, StarDental Part #256719, through entire tip starting at
opening in front.
CLEANING Titan SW or S Water Tube
On a regular basis or if a water tube blockage occurs,
remove the tip, and on the Titan SW also disconnect the
swivel. Push the cleaning wire through the internal water
tube using needle nose pliers at the front on both the Titan
SW and S as shown. Push small lengths of 3 to 4 mm at a
time until the wire emerges at the opposite end. Pull wire
completely out.
Cleaning Wire
Opening
Front - Titan K and SW
StarDental® Instruments
Question and Answer IconNeed help?

Do you have a question about the StarDental Titan S and is the answer not in the manual?

Summary

O-Rings & Rear Seal Replacement

Swivel O-Rings

Periodically replace swivel O-rings for proper scaler function.

Front Suspension O-Rings

Replace front suspension O-rings after two years to fix play between shaft and handle.

Titan S;SW Rear Seal & Swivel O-Rings

Replace rear seals annually or as needed for optimal performance.

2;3-line for Titan S;SW

Procedure for replacing the 2/3-line rear seal and swivel O-rings.

Cleaning Wire

CLEANING the Tips

Clean tips weekly by running cleaning wire through the entire tip.

CLEANING Titan SW or S Water Tube

Push cleaning wire through internal water tube to clear blockages.

CLEANING the Swivel

Daily cleaning of the swivel and mating components to ensure peak performance.

Water Tube Gasket Replacement

2;3 Line

Steps to replace the water tube gasket on 2/3-line scaler assemblies.

4 Line

Steps to replace the O-ring on the water tube assembly for 4-line scalers.

StarDental Titan S Specifications

General IconGeneral
CategoryDental equipment
AutoclavableYes
Fiber Optic IlluminationYes
Chuck TypePush Button
Noise Level65 dB
Connection4-hole
Bur TypeStandard
SpraySingle

Related product manuals