If the Current Output Board is to be fitted in the unit by the user, it is necessary to
use this entry to calibrate the option.
Connect up the Current output board to the Sensor input board as shown below.
Press ENTER, to commence automatic
calibration.
Note if the display reads either 'siglo' or
'sighi' recalibrate the output card again.
On completion ESCAPE back to the Process
Variable display mode. This stores the
calibration information in non-volatile
memory.
4.2.5.4 CLvoP1(3) Calibrate Bridge excitation
board
CALIBRATION MENU
Page 66