Users should be trained, familiarize with instructions, and follow cautions.
Detailed description of controls, displays, connectors, and their functions.
Risk of explosion if used with flammable anesthetics.
Risk of electrical shock, do not remove cover.
Read and follow instructions carefully; understand Warning, Caution, Note.
Observe instructions and interface specifications for combined devices.
Risk of electric shock; unauthorized opening voids warranty.
Patient leakage currents can be additive; use Type CF accessories.
Avoid liquid entry into the unit; do not store liquids on or above it.
Use as accessory with rigid/flexible endoscopes; with compatible camera heads.
Combinations are safe if identified as such or if specifications permit.
Check outer surfaces for rough surfaces, sharp edges or protrusions.
Operate unit only with the voltage stated on the identification plate.
May only be used by physicians and medical assistants who are qualified.
Read manual thoroughly before use; familiarize with safety sections.
Always unplug the unit before performing maintenance.
Flammable gases may be present; dissipate them before use.
Use only fuses of the same rating as specified.
Use only in medical rooms with compliant electrical systems.
Not for use in hazardous zones; avoid flammable gases.
User is responsible for ensuring safe operation and proper condition before use.
Carefully unpack, check for missing items, file complaints for shipping damage.
Connect accessories according to national, state, and local electrical codes.
Connect power cord to receptacle; do not connect/disconnect in hazardous zones.
Connect DV video output to recorder via DV cable.
Optimal image quality via direct signal transmission; monitor monitoring.
Ensure camera head connector is clean and dry before plugging into CCU.
Diagram showing connections for CCU, VCR, printer, monitor, HD display, light source.
Switch on CCU and monitor; color bars will appear for 2 seconds.
Lists the IMAGE1 HUB™ HD CCU and included accessories.
Place CCU on a level surface, ensure sufficient distance from other equipment.
Connect CCU to video monitor via DVI, RGB, S-Video, or Composite output.
Connect accessory devices (e.g., printer, VCR) to ACC1/ACC2 ports.
Keyboard required for patient data entry and camera head programming.
Ensure camera head connector is clean and dry before plugging into CCU.
Connect endoscopes and light cables to camera heads.
Connect the potential equalization connector with a ground line.
Connect SCB cable with protection device to SCB sockets.