12 Installation & Operation Manual for the
Active Vibration Cancellation (AVC) System
7) Display the integral constant of the temperature control loop
a) Command: KI<CR>
b) Returns the integral constant of the temperature control loop.
8) Set the integral constant of the temperature control loop
a) Command: KI=<VAL><CR>
b) This command is User locked.
c) <VAL> is the user defined integral constant of the temperature control loop.
9) Display the proportional constant of the temperature control loop
a) Command: KP<CR>
b) Returns the proportional constant of the temperature control loop.
10) Set the proportional constant of the temperature control loop
a) Command: KP=<VAL><CR>
b) This command is User locked.
c) <VAL> is the user defined proportional constant of the temperature control loop.
11) Display the current status of the LCD
a) Command: LCD<CR>
b) Returns the current status of the LCD.
12) Set the state of the LCD
a) Command: LCD=<VAL><CR>
b) This command is User locked.
LCD states:
ON LCD is regularly updated.
OFF LCD is not updated.
KI
001.00000
KI=0.5
000.50000
KP
001.00000
KP=5
005.00000
LCD
OFF
LCD=ON
ON