Installation & Operation Manual for the 15
Active Vibration Cancellation (AVC) System
23) Display the temperature sensor reading in kelvin
a) Command: TC<CR>
b) Returns the cold head temperature in kelvin.
24) Display the reject temperature in Celsius
a) Command: TEMP<SP>RJ<CR>
b) Returns the current reject temperature measured at the base of the fins on the cooler. Reject temperature is displayed
in Celsius.
25) Display target temperature in kelvin
a) Command: TTARGET<CR>
b) Returns the target temperature of the controller in kelvin when in temperature control mode (default mode).
26) Set target temperature
a) Command: TTARGET=<VAL><CR>
b) This command is User locked.
c) Set the temperature the cryocooler will try to attain (to ±0.1 K) when in temperature control mode.
d) <VAL> is the value of the desired target temperature in kelvin.
27) Display current controller code version
a) Command: VERSION<CR>
b) Returns the current controller coder version.
TEMP RJ
025.30
TC
295.21
TTARGET
077.00
TTARGET=77
077.00
VERSION
1.0.4