Deluxe Nasal Pillow CPAP Mask
Item# CM036
Intended Use
The nasal pillow mask is intended to provide an interface for Continuous Positive Airway Pressure (CPAP) or bilevel system.
The mask to be used by adult patients (>66lb/30kg) for whom positive airway pressure has been prescribed. The mask is for
single-patient re-use in the home environment and multi-patient re-use in the hospital/institutional environment.

Caution
US federal law restricts this device to sale by or on the order of a physician.
Excluding patient population
This mask may not be suitable for patients with the following conditions: impaired cardiac sphincter function, excessive
reflux, impaired cough reflex and hiatal hernia. It should not be used if the patient is unresponsive, obtunded, unable to
remove the mask, or uncooperative.

Warnings
This mask is designed for use with CPAP or bilevel systems recommended by your health care professional or respiratory
therapist. Do not wear this mask unless the CPAP or bilevel system is turned on and operating properly. Do not block or try
to seal the exhalation port.
CPAP systems are intended to be used with special masks with connectors which have vent holes to allow continuous ow of air out of the mask. When the
device is turned on and functioning properly, new air from the device ushes the exhaled air out through the attached mask exhalation port. However, when
the device is not operating, enough fresh air will not be provided through the mask, and exhaled air may be re-breathed. At low pressures, the airow through
the exhalation port may not be enough to clear all of the exhaled gas (CO2) from the mask. You may breathe in some of the air that you have exhaled.
If oxygen is used with the device, the oxygen flow must be turned off when the device is not operating.
When the device is not in operation, and the oxygen ow is left on, oxygen delivered into the ventilator tubing may accumulate within the device enclosure.
Oxygen accumulated in the device enclosure will create a risk of re.
Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of an open ame.
At a xed ow rate of supplemental oxygen ow, the inhaled oxygen concentration will vary, depending on the pressure settings, patient breathing pattern,
nasal interface selection, and the leak rate.
Some users may experience skin redness. If this happens, contact your health care professional.
Cleaning the Mask
The cushion, nasal interface, and tubing should be cleaned daily or as needed.
Hand wash in warm water with a mild liquid dishwashing detergent. Do not use bleach, alcohol, or cleaning solutions
containing alcohol. Do not use cleaners containing conditioners or moisturizers.
Rinse thoroughly. Pat with a towel and air dry completely before use
Inspect all parts for damage or wear; replace if necessary.
Cleaning the Headgear
Hand wash the headgear weekly or as needed. Lay the headgear flat or line dry.
Disposal
The nasal pillow mask does not contain any hazardous substances and may be disposed of with your normal
household refuse.
latexfree
Manufactured for
Sunset Healthcare Solutions
180 N Michigan Ave Ste 2000
Chicago, IL 60601
SunsetHCS.com
DO NOT PRINT THIS LAYER.
100%
1 Color
1
PH 2.23.17
A4 (210mm x 297mm)
res036-ifu.pdf
When printing with default
Acrobat autoscale settings.
Instruction Manual
Compressor Nebulizer
NEB100
NEB100-R
NEB100-IFU V 4.00
2019-11-07
EN English
ES Español
Question and Answer IconNeed help?

Do you have a question about the Sunset Healthcare NEB100-R and is the answer not in the manual?

Sunset Healthcare NEB100-R Specifications

General IconGeneral
Product TypeCompressor Nebulizer
ModelNEB100-R
Compressor TypePiston
Power SourceAC
Nebulization Rate≥0.2ml/min
Particle SizeMMAD ≤5 μm
Air Flow6-8 L/min
Included AccessoriesMouthpiece, Tubing, Filter

Summary

Instructions for Use

Caution

Federal law restricts this device to sale by or on the order of a physician.

Warranty Information

Details on the limited 3-year warranty for the compressor unit and return policy.

Using the T-Piece Kit

Troubleshooting Guide

Steps to resolve issues like the device not turning on or weak nebulization.

Device Maintenance

Cleaning Procedures

Instructions for cleaning the compressor unit and its accessories.

Nebulizer Kit Replacement

Guidance on when and how to replace disposable or reusable nebulizer kits.

Filter Replacement

Information on replacing the nebulizer filter after approximately 30 hours of use.