s
Monitor Operating Keys
B
Alarm Silence
Use this key to silence the alarms and to cancel the alarm silence.
Menu / Enter Key
Press this key to enter the menu system, and to advance to the next
menu selection.
Cancel / Exit Key
Press this key to exit the menu system.
n
Up and
o
Down Arrows
Use the up and down arrow keys to adjust pulse beep volume during
normal operation. In a menu, use these keys to adjust the selection.
x
ON/OFF Key
Momentarily press this key when the device is OFF to turn the monitor ON.
Press and hold this key when the device is OFF to clear the patient trend data.
Silencing Alarms
Audible alarms can be silenced for two (2) minutes.1.
To silence the alarm tone, press the 2. Alarm Silence key (
B
).
The Alarm Silence indicator will light solid for the rst minute and 45 seconds, 3.
then ash the last 15 seconds.
To cancel the 2-minute silence, press the 4. Alarm Silence key (
B
).
NOTE: The Alarm Silence key will not aect visual alarm indicators or pulse tone.
NOTE: Permanent Alarm Silence is not available.
Changing the Alarm Limits
To select alarm limit defaults by patient type, press the 1. Menu key (
) once to display the current patient type
setting. Pt will be displayed as well as the current setting: Horse (KrS), Dog (d09), Cat ([At), Factory (FA[), or
Custom ([Us).
Use the 2. Up and Down keys (
n
o
)to select the desired patient type.
Press the 3. Menu key (
) again to display the desired alarm limit values for that patient type.
Use the 4. Up and Down keys (
n
o
) to increase or decrease the setting.
Press the 5. Menu key (
) to advance to the next alarm limit.
To exit, press the 6. Exit key (
) or wait 20 seconds until the patient reading is displayed.
Page 1 of 2
6 For use by or on the order of a licensed veterinarian.
f Attention, see operation manual for additional instructions for use.
Quick Reference Guide
V1030
5
4
3
2
1
4
Surgivet® V1030 Hand Held Pulse Oximeter
Simplied Operation Guide
B
B
1
5
4
3
2
4
Question and Answer IconNeed help?

Do you have a question about the SurgiVet V1030 and is the answer not in the manual?

SurgiVet V1030 Specifications

General IconGeneral
BrandSurgiVet
ModelV1030
CategoryMedical Equipment
LanguageEnglish

Summary

Monitor Operating Keys

Alarm Silence Key

Use this key to silence the alarms and to cancel the alarm silence.

Menu / Enter Key

Press this key to enter the menu system and advance to the next menu selection.

Cancel / Exit Key

Press this key to exit the menu system.

Navigation Arrows

Use the up and down arrow keys to adjust pulse beep volume or menu selections.

On/Off Key

Momentarily press to turn the monitor ON. Press and hold to clear patient trend data.

Silencing Alarms

Changing Alarm Limits

Pulse Volume Control

Entering Setup Menu

Setting Configuration Options

Setting Averaging

Select desired SpO2 averaging (1-4) and advance to the next menu or exit.

Setting Sensitivity

Turn the High Sensitivity setting On or OFF.

Display Brightness

Select desired display brightness, ranging from 1-10.

Setting the Date

Select the year, month, and day for the device.

Setting the Time

Select the hour and minute for the device.