Maintenance
6 monthly
Clean the Oxytrace G membrane with a soft tissue.
Clean the Faraday electrode (if installed)
Yearly
If necessary, replace filling electrolyte. If the sensor is exposed to air frequently and
during long time intervals, the electrolyte and membrane may have to be changed
earlier.
Clean flow cell and flow meter, if dirty
Every 2
nd
year
Replace Swansensor Oxytrace G membrane by using a new, pre-filled sensor cap.
Oxytrace
Maintenance
Video
Oxytrace
Verification
Video
Oxytrace
Calibration
Video
How to remove Oxytrace Sensor
A Fixing sleeve
B Sensor
C Flow cell
Unscrew the fixing sleeve [A].
Remove the sensor from the flow cell.
Note: A vacuum may form when the sensor is pulled out. Do not use force!
The sensor can be easily removed if the flow control valve is slightly opened.
Reorder
codes
Description
Image
Sensor Cap Oxytrace G
Error
Messages
To view active error messages, press ENTER to open menu system, select
MESSAGES -> PENDING ERRORS. Consult the manual for a full list of error
messages.
HINT: With any Swan system, press ENTER 3 times from the home screen to bring
up the PENDING ERRORS list.