EasyManua.ls Logo

Tegam 840A - Page 19

Default Icon
47 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
• To correct a numeric entry while in the NUM mode, hit the
CHANGE/ENTER key twice. This will return number entry to
the beginning of a new number.
• To return the calibrator output to 0° while in the OPR mode, hit
the CLEAR key.
• The unit can be used as a convenient °F to °C or °C to °F cal-
culator.
For example, to convert 77°F to a Celsius reading, go to the CALIB OPR
mode, enter the reading to be converted, and change scale.
eg.°F
CHANGE/ENTER
77
CHANGE/ENTER
°C (Display reads 25°C)
7. CALIBRATOR MODE OPERATION (RTD, Thermistor & Ohms)
Set-Up:
1. Connect measurement equipment under test (eg. RTD thermometer,
ohmmeter, etc.) to the instrument as shown in Figures 8, 9, or 10.
The same performance tradeoffs apply to 2, 3 and 4-wire simulations as
to 2, 3, and 4-wire measurements. Refer to RTD, Thermistor and Ohms
Meter-Mode section above for wiring guidelines.
2. Set instrument to CALIBRATOR mode.
3. Use the sensor-select key to set the instrument to the appropriate func-
tion and range.
Figure 8. 2-Wire Resistance Simulation
18

Related product manuals