8. STORAGE AND RECALL OF CALIBRATOR SETTINGS
While operating as a calibrator, the unit can store and recall up to 10 output
settings (each with independent parameters, eg. °F/°C, sensor type, and res-
olution) in random access memory.
To Store Outputs:
1. While in the CALIB OPR mode, depress the STORE key. Note that the
OPR annunciator is replaced by STO.
2. Next depress one of the memory location numbers (keys 0, 1, 2, .... 9).
The corresponding memory annunciator turns on, and the calibrator
returns to the OPR mode.
eg. STORE 2
NOTES:
• The enabled memory annunciator shows that the output set-up
has been saved.
• When a set up is stored in a location already in use, the new
set-up replaces the old.
• Saved set-ups are retained during power-off.
• To store a reading of 0° just hit the CLEAR while in the CALIB
OPR mode. Then depress STORE and a location number.
To Recall Outputs:
1. While in the CALIB OPR mode, set-ups saved earlier in memory can be
recalled to program the calibrator output. First depress the RECALL key
to recall a set-up.
Note that the RCL annunciator turns on, OPR turns off, and all memory
locations annunciators are blanked.
2. Next hit the key with the number corresponding to the desired memory
location. The corresponding annunciator turns on, and the LCD updates
to show the new calibrator setting.
eg.RECALL 8
3. Several stored set-ups can be randomly recalled by a key sequence such
as:
eg.RECALL 6
RECALL 3
RECALL 9
etc.
20