POWER SUPPLY ISOLATION PROCEDURE
Each regulated supply has numerous feed points to on that supply line. By lifting jumpers or other components
external loads through the instrument. Diagram
8,
power in the supply line farther down the line, the circuit in which
distribution, is used in conjunction with the schematic a short exists may be located.
diagrams to determine the service jumper or component
that may be lifted to isolate loads from the power supply.
Always set the POWER switch to OFF before soldering
or unsoldering service jumpers or other components and
If a supply comes up after lifting one of the isolating before attempting to measure component resistance
jumpers, it is very probable that short exists in the circuitry values.