EasyManua.ls Logo

TEK 2225 - Page 69

TEK 2225
215 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Adjustment Procedure-2225 Service
1
Table
5-3
Deflection Accuracy Limits
6.
Check Input Coupling
a. Set the AC-GND-DC switches (both channels)
to GND.
b. Position the trace on the center horizontal
graticule line using the CH 2 POSITION control.
c. Change the generator output to 50
mV
d. Set the CH 2 AC-GND-DC switch to AC.
e. CHECK-That the display is centered about the
center horizontal graticule line.
f. Set the CH
2
AC-GND-DC switch to DC.
g. CHECK-That the display is ground referenced
on the center horizontal graticule line.
h. Move the test-signal cable from the CH 2
OR
Y
input connector to the CH
1
OR
X
input connector.
i.
Set the Vertical MODE switch to CH
1.
j.
Repeat Parts b through
g
using the Channel
1
controls.
7.
Check Position Range
a.
SET:
VOLTSIDIV (both)
10
mV
AC-GND-DC (both) AC
SECIDIV
0.2 ms Trigger
Trigger COUPLING
AC
b. Set the calibration generator for
0.1
V.
c.
Adjust the CH
1
VOLTSIDIV Variable control to
produce a 5.25-division display.
d. Set the CH
1
VOLTSIDIV to
5
mV.
I
e. Set the calibration generator to produce a
0.2 V signal.
I
f. CHECK-The bottom and top of the trace may
I
be positioned above and below the center horizontal
graticule line by rotating the CH
1
POSITION control
fully clockwise and counterclockwise respectively.
g. Move the cable from the CH
1
OR
X
input con-
nector to the CH 2
OR
Y
input connector.
h. Set the Vertical MODE switch to CH 2.
I
i.
Repeat Parts b through f using the Channel 2
controls.
j.
Disconnect the test equipment from the instru-
ment.
8.
Adjust
XllXlO
Balance
a. SET:
Vertical MODE CH
1
AC-GND-DC (both) GND
VOLTSIDIV Variable (both) CAL detent
b. Position the trace on the center horizontal
graticule line using the CH
1
POSITION control.
c. Set CH
1
VOLTSIDIV Variable knob to
XI0
(pull
CAL knob out).
d. ADJUST-XI
0
BAL
(R107)
to position the trace
I
on the center horizontal graticule line.
e. Set CH
1
VOLTSIDIV Variable knob to
X1
(push
CAL knob in).
f.
Repeat Parts b through e until there is no trace
shift between
XI
and
XI0
positions.
Scam
by
ARTEK
MEDLa
=>
REV
APR
1990