EasyManua.ls Logo

TEK 2225 - Page 70

TEK 2225
215 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Adjustment Procedure-2225 Service
g. Set Vertical MODE to CH 2.
h. Repeat Parts b through f for CH 2, using the
Channel 2
XI 0 BAL adjust (R157) instead of R107 in
Part d.
i. Return both VOLTSIDIV Variable controls to their
CAL and
XI positions.
9.
Adjust Attenuator Compensation
a. SET:
VOLTSIDIV (both) 5 mV
Vertical Magnification (both) XI (CAL
knobs in
AC-GND-DC (both) DC
b. Connect the high-amplitude, square-wave
output from the calibration generator via a
504 BNC
termination, a probe-tip-to-BNC adapter, and the
10X probe to the CH 2
OR
Y
input connector.
c. Set the generator to produce a 1-kHz,
five-
division display and compensate the probe using the
probe compensation adjustment (see the probe
instruction manual).
d. Set the CH 2
VOLTSIDIV switch to 10 mV.
Figure
5-1
.Attenuator trimmer adjustments.
I
FRONT
OF
e. Replace the probe and probe-tip-to-BNC
adapter with a
504 BNC coaxial cable and 504 BNC
I. Repeat Parts e through
j
except adjust the
"2"
termination.
and
"2N" trimmers in Parts g and
j
respectively.
-
f. Set the generator to produce a five-division
display.
INSTRUMENT
m. Set the CH 2 VOLTSIDIV switch to 50 mV.
-
6299-32A
n. Repeat Parts e through j except adjust the
"
3"
g. ADJUST-Trimmer 1 for flattest response on
and
"3N"
trimmers
in
Parts
and
respectively.
the square wave signal. See figure 5-1 for location of
the trimmers.
o. Set the CH 2 VOLTSIDIV switch to
.5 V.
h. Replace the
504 BNC coaxial cable and 504
BNC termination with the probe and probe-tip-to-
p. Repeat Parts e through
j
except adjust the "4"
BNC adapter.
and
"4N" trimmers in Parts g and j respectively.
i.
Set the generator to produce a five-division
q.
Set the Vertical MODE switch to CH 1.
square wave.
r. Repeat Parts b through p for the Channel 1
At-
j. ADJUST-Trimmer 1 N for flattest response on
tenuators.
square wave.
s. Disconnect the test equipment from the in-
k. Set the CH 2
VOLTSIDIV switch to 20 mV.
strument.
REV
APR
1990
scans
by
ARTEK
MEDLQ
=>