EasyManua.ls Logo

Tektronix 502A - Page 16

Tektronix 502A
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating
Instructions
Type
502A
If
the
trace
shifts
slightly
between
the
different
ranges
of
the
SENSITIVITY
Control
this
may
be
cured
by
a
very
slight
readjustment
of
the
DC
BAL
Control.
The
following
pro
cedure
may
be
used
to
correct
for
this
slight
trace
shift.
Set
the
free-running
trace
to
a
convenient
horizontal
graticule
line
with
the
SENSITIVITY
Control
in
the
.2
VOLTS
PER
CM
position.
Rotate
the
SENSITIVITY
Control
to
the
.1
mVOLTS
PER
CM
position
and
bring
the
trace
back
to
the
original
horizontal
graticule
line
with
the
DC
BAL
Control.
NOTE
The
fast
area"
with
the
FINDER
Button
depressed
can
only
be
easily
seen
on
the
higher
sensitivity
settings
of
the
SENSITIVITY
Control.
Single
Sweep
Single
sweep
is
used
to
record
waveforms
which
are
not
repetitive
or
waveforms
which
jitter
too
much
for
normal
triggered-display
photography.
To
use
the
single
sweep
feature
of
the
Type
502A,
set
the
MODE
switch
to
NORMAL.
If
the
waveform
is
repetitive
it
should
be
connected
to
the
input
of
the
oscilloscope.
If
the
waveform
is
not
repetitive
then
an
AMPLITUDE
CALI
BRATOR
signal
of
the
same
amplitude
as
that
of
the
expected
signal
should
be
used.
Adjust
the
trigger
control
to
obtain
a
stable
display
of
the
input
signal.
In
the
case
of
a
waveform
that
jitters,
adjust
the
control
to
obtain
the
most
stable
display.
If
a
calibrator
waveform
was
used
to
adjust
the
trigger
control,
disconnect
the
calibrator
waveform
from
the
input
to
the
oscilloscope
and
connect
the
non-repetitive
waveform
source
to
the
oscilloscope.
Push
the
MODE
switch
to
the
RE
SET
position.
The
READY
light
should
now
be
lit.
The
Type
502A
will
now
sweep
once,
on
the
first
trigger
received
after
the
MODE
switch
is
released;
further
signals
will
not
cause
the
instrument
to
sweep.
If
the
waveform
is
repetitive
leave
the
signal
connected
to
the
input
connector
of
the
oscilloscope.
When
a
sweep
is
desired,
push
the
MODE
switch
to
RESET
and
let
it
return
to
the
SINGLE
SWEEP
position.
Upon
the
return
of
the
MODE
switch
to
the
SINGLE
SWEEP
position,
the
sweep
will
sweep
once.
If
another
sweep
is
desired
the
MODE
switch
must
again
be
pushed
to
the
RESET
position
and
returned
to
the
SINGLE
SWEEP
position.
It
may
be
quite
convenient
to
use
the
READY
light
of
the
Single
Sweep
circuit
as
an
indicator
for
photography.
The
READY
light
will
light
when
the
circuit
is
awaiting
a
trigger
and
will
stay
lit
until
the
sweep
has
been
completed.
At
the
end
of
the
sweep
the
READY
light
will
go
out
and
stay
out
until
the
circuit
is
reset
for
another
sweep
with
the
MODE
switch.
INTENSITY
BALANCE
Control
The
INTENSITY
BALANCE
Control
is
used
to
adjust
the
in
tensity
of
both
vertical
amplifiers
to
the
same
level.
With
both
traces
adjusted
to
the
same
level,
each
will
have
the
same
relative
writing
rate,
hence
each
trace
will
show
the
same
amount
on
a
photograph.
It
may
be
necessary
to
readjust
the
INTENSITY
BALANCE
Control
each
time
the
setting
of
the
INTENSITY
Control
is
changed.
This
is
due
to
the
characteristics
of
a
mono
accelerator
crt
in
which
the
triode
characteristics
for
each
crt
gun
are
slightly
different.
AUXILIARY
FUNCTION
AMPLITUDE
CALIBRATOR
The
AMPLITUDE
CALIBRATOR
is
a
source
of
accurately
calibrated
square
waves
at
a
frequency
of
approximately
1
kilocycle
(
+
or
30%).
The
output
amplitude
is
accurate
within
3%
of
the
AMPLITUDE
CALIBRATOR
switch
settings
for
no
load
conditions.
The
primary
function
of
the
calibrator
is
to
provide
a
convenient
method
for
checking
the
calibra
tion
of
the
vertical-deflection
system
and
for
adjusting
the
probes.
It
can
be
used
for
other
purposes,
however,
if
it
is
kept
in
mind
that
the
output
impedance
of
the
cali
brator
varies
with
the
output-voltage
setting.
Trace-brightness
(intensity)
modulation
To
couple
markers
or
other
signals
into
the
crt
cathodes
for
trace-brightness
modulation,
disconnect
the
ground
strap
at
the
rear
of
the
instrument
and
apply
the
signal
between
the
CRT
CATHODE
binding
post
and
GND
.
This
will
simultan
eously
apply
the
signals
to
the
upper
and
lower
beam
cathodes.
At
normal
brightness,
positive
signals
of
about
25
volts
will
cut
the
beams
off.
Always
replace
the
ground
strap
when
you
are
not
intensity
modulating
the
beams.
Graticule
illumination
The
graticule
lighting
control,
labeled
POWER
AND
SCALE
ILLUM,
can
be
ad|usted
to
suit
the
lighting
conditions
of
the
room.
A
light
filter
is
supplied
with
the
instrument
which
can
be
used
for
increased
contrast.
This
filter
should
be
mounted
next
to
the
crt
face
so
it
does
not
block
the
light
from
the
graticule
lines.
The
graticule
of
the
Type
502A
Oscilloscope
can
be
illuminated
so'
that
it
appears
to
have
either
red
or
white
graticule
markings.
The
markings
can
be
quickly
changed
from
white
to
red
or
from
red
to
white
by
removing
the
graticule
cover
and
inverting
the
graticule.
As
a
general
rule,
white
graticule
lines
are
superior
to
red
for
photo
graphic
purposes.
Direct
connection
to
crt
plates
In
some
applications,
it
may
be
desirable
to
connect
a
signal
directly
to
one
or
more
sets
of
crt
deflection
plates
bypassing
the
internal
oscilloscope
amplifiers.
This
can
be
done
in
the
Type
502A
if
certain
precautions
are
observed.
One
of
these
precautions
is
to
m
a
intain
the
aver
age
d
c
voltage
on
the
deflection
plates
at
approximotely
+225
volts.
If
the
average
voltage
is
not
maintained
at
approximately
+225
volts,
the
crt
display
may
be
defocused.
For
many
applications,
ac
coupling
of
the
signal
to
the
deflection
plates
is
permissible.
It
has
the
advantage
of
allowing
the
use
of
front
panel
controls
to
position
the
display
and
permits
the
use
of
signals
not
having
the
re
quired
average
voltage
level.
A
diagram
of
this
method
of
2-10
®i

Related product manuals