EasyManua.ls Logo

Tektronix AFG3021B - Page 60

Tektronix AFG3021B
113 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating basics
Maximum leadin
gedgetime. This value is the minimum of the three in each
instance.
If runMode = Co
ntinuous:
Temp1=0.8*2.0*width–tEdge;
Temp2=(period–width)*0.8*2.0–tEdge;
Temp3 = 0.625 * period.
Else:
Temp1=0.8*2.0*width–tEdge;
Temp2 = ( p
eriod leadDelay width ) * 0.8 * 2.0 tEdge;
Temp3 = 0.625 * period.
Maximum trailing edge time. This value is the minimum of the three in each
instance.
If runMode = Continuous:
Temp1
=0.8*2.0*width–lEdge;
Temp2=(period–width)*0.8*2.0–lEdge;
Temp3 = 0.625 * period.
Else:
Temp1=0.8*2.0*width–lEdge;
Temp2=(period–leadDelay–width)*0.8*2.0–lEdge;
Tem
p3 = 0.625 * p eriod.
46 AFG3000 and AFG3000C Series User Manual

Table of Contents

Related product manuals