Getting started
1. Push the front-panel Utility
button.
2. Push the -more
- bezel
button.
3. Push the
Diagnostics
/Calibration
bezel button.
4. To execute t
he instrument
diagnostics, push the
Execute Diagnostics
bezel butto
n.
To execute self calibration,
push the Execute
Calibrati
on bezel button.
5. If Diagnostics completes
without any errors, the
message
PASSED” is
displayed.
CAUTION
. Do not power off the instrument while executing self calibration. If the
power is turned off during self calibration, data stored in the internal memory
may be lost.
Quick Tips
Before executing self calibration, ensure that the ambient temperature is
between +20 °C and +30 °C (+68 °F to +86 °F). Allow a 20 minute warm-up
period before executing self calibration.
Disconnect all the cables from the instrument when you perform self test or
self calibration.
Perform self calibration at least once a year to maintain DC accuracy. It is
recommended that the self calibration should be performed along with a
periodic check.
10 AFG3000 and AFG3000C Series User Manual

Table of Contents

Other manuals for Tektronix AFG3102

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix AFG3102 and is the answer not in the manual?

Tektronix AFG3102 Specifications

General IconGeneral
BrandTektronix
ModelAFG3102
CategoryPortable Generator
LanguageEnglish

Related product manuals