Model 7700 Multiplexer Module Instructions for use with 2750 
 
077182400  July 2023  5 
 
To prevent electric shock that could result in serious injury or death, comply with these safety 
precautions:   
  Before making or breaking any connections to the switching card, make sure the 
instrument is turned off and power is removed from all external circuitry. 
  Do not connect signals that will exceed the maximum specifications of any installed 
switching card. 
Dangerous arcs of an explosive nature in a high-energy circuit can cause severe personal 
injury or death if contacted. If the instrument is connected to a high-energy circuit when set to 
a current range, low-resistance range, or any other low-impedance range, the circuit is 
virtually shorted. Dangerous arcing can result even when the multimeter is set to a voltage 
range if the minimum voltage spacing is reduced in the external connections. For details about 
how to safely make high-energy measurements, see “High-energy circuit safety precautions” 
in the Reference Manual for the instrument. 
 
Do not exceed the maximum specifications for the 7700. Refer to the specifications provided in 
the datasheet. Failure to recognize and observe normal safety precautions could result in 
personal injury or death. 
The following information describes how to make connections to the switching module and define the channel 
designations. A connection log is provided that you can use to record your connections. See Connection log
 
(on page 10). 
This section describes how to make connections to the terminal screws in the module. You can make: 
  Current connections, provided through two protected channels (channels 21 and 22) 
  INPUT connections 
  SENSE (4-wire resistance) connections 
  AMP and LO common connections to the instrument 
Channel 23 (2W/4W Configuration), Channel 24 (Sense Isolation), and Channel 25 (Input Isolation) are normally 
automatically configured by the 2750 for system channel operation. However, by using multiple channel 
operation, channels can be individually controlled. 
Connect 4-wire sense leads using channels 11 to 20. To disconnect channels 11 to 20 from channels 
1 to 10, send: 
ROUT:MULT:CLOS (@123) 
Note opposite logic.