Commands Listed in Alphabetical Order
Group
Bus
Syntax
BUS:B<x>:SPI:IDLETime <NR3>
BUS:B<x>:SPI:IDLETime?
Arguments
<NR3> species the idle time in seconds.
Examples
BUS:B1:SPI:IDLETime? might return BUS:B1:SPI
:IDLETime 5.0000E-6
indicating that the idle time is 5 μs.
BUS:B<x>:SPI{:SELect|:SS}:POLARity
Sets or returns the SPI SS polarity for bus <x>, where x is the bus number.
Conditions
This command requires a DPO2EMBD application module.
Group
Bus
Syntax
BUS:B<x>:SPI{:SELect|:SS} :POLARity {LOW|HIGH}
BUS:B<x>:SPI{:SELect|:SS} :POLARity?
Arguments
LOW species an active low polarity.
HIGH species an active high polarity.
BUS:B<x>:SPI{:SELect|:SS}:SOUrce
Sets or returns the SPI SS source for bus <x>, where x is the bus number.
Conditions
This command requires a DPO2EMBD application module.
Group
Bus
Syntax
BUS:B<x>:SPI{:SELect|:SS} :SOUrce {CH1|CH2|CH3|CH4|
D0|D1|D2|D3|D4|D5|D6|D7|D 8|D9|D10|D11|D12|D13|D14|D15}
BUS:B<x>:SPI{:SELect|:SS} :SOUrce?
2-90 MSO2000B, DPO2000B, MSO2000 and DPO2000 Series O scilloscopes Programmer Manual