Commands Listed in Alphabetical Order
MEASUREMENT:M
EAS3:DELAY:DIRECTION?
might return
:MEASUREMENT:MEAS3:DELAY:D IRECTION BACKWARDS indicating
that the current search direction is backwards.
MEASUrement
:MEAS<x>:DELay:EDGE<x>
Sets or returns the slope of the edge used for the delay "from" or "to" waveform
when taking an immediate delay measurement. The waveform is specied by
MEASUreme
nt:MEAS<x>:SOURCE[1].
Group
Measurement
Syntax
MEASUrement:MEAS<x>:DELay: EDGE<x> {FALL|RISe}
MEASUrement:MEAS<x>:DELay: EDGE<x>?
Arguments
<x> species which waveform to use, wher
e <x> = 1 is the "from" waveform, and
<x> = 2 is the "to" waveform.
FALL species the falling edge.
RISe species the rising edge.
Examples
MEASUREMENT:MEAS1:DELAY:ED GE1 RISE species that the "from" waveform
rising edge be used for the immediate delay measurement.
MEASUREMENT:MEAS1:DELAY:ED GE1? returns either RISE or FALL.
MEASUrement:MEAS<x>:MAXimum? (Query Only)
Returns the maximum value found for this measurement since the last statistical
reset. Measurements are specied by x, which ranges from 1 through 4.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MAXimu m?
Examples
MEASUREMENT:MEAS3:MAXIMUM? might return
:MEASUREMENT:MEAS3:MAXIMUM 4.18E-9.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-199