Type &
min. size
Function Type of
user info
stored
Method of
modificatio
n
Data Input
method
Location User
accessible
To clear To sanitize
EEPROM
≥2 Kbit
Stores
factory
data,
maintenanc
e data, and
user
password
User
settings
User
password is
settable
using PI
commands
Indirect Factory
operations
and
programmat
ic
commands
Acquisition
board
Yes Overwrite
user
password.
(see
Overwriting
the user
password
on page 8
.) Clearing
entire
memory
device
would
disable
instrument
functionality.
Overwrite
user
password.
(see
Overwriting
the user
password
on page 8
.) Sanitizing
entire
memory
device
would
disable
instrument
functionality.
EEPROM
≥2 Kbit
Holds AFG
calibration
data
None Indirect Factory
operations
AFG riser
board
No Not
applicable,
does not
contain user
data or
settings.
Clearing
would
disable
instrument
functionality.
Not
applicable,
does not
contain user
data or
settings.
Sanitizing
would
disable
instrument
functionality.
EEPROM
≥64 Kbit
Holds the
front panel
USB
configuratio
n
None None Factory
operations
Front panel
LED board
No Not
applicable,
does not
contain user
data or
settings.
Clearing
would
disable
instrument
functionality.
Not
applicable,
does not
contain user
data or
settings.
Sanitizing
would
disable
instrument
functionality.
Table continued…