Appendix A: Specifications
26
PS280 & PS283 User Manual
Table 3: Operational Characteristics (Cont.)
Characteristic Measurement
Line Regulation (CC)
0.2% +3 mA
Load Regulation (5 V)
0.2%
Load Regulation (CV)
0.01% +3 mV (rating current 3A)
0.01% +5 mV (rating current >3 A)
300 mV (0–60 V single series tracking supply)
Load Regulation (CC)
0.2% +3 mA
Ripple/Noise (5 V)
2 mV rms
Ripple/Noise (CV)
1 mV rms, 5 Hz–1 MHz
Ripple (CC)
3 mA rms
Temperature Coefficient (CV)
300 ppm/_C
Recovery Time (CV)
100 s (time to recover after a 50% load change with
0.5 A minimum)
Tracking Error (Slave)
0.5% +10 mV of the master supply
Indicator Two 3 1/2 digit 0.5 in LED panel display meter
Meter Indicators 0–30 VDC ±(0.5% of reading + 2 digits)
0–2 A ±(0.5% of reading + 2 digits)
Insulation
(Chassis-to-Terminal)
20 M at DC 500 V
Insulation
(Chassis-to-AC Cord))
30 M at DC 500 V