Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–217
"hello";STATE 1;BOX 74,84,475,135;:LOCK NONE; :HARDCOPY:FORMAT
EPSIMAGE;PORT GPIB;LAYOUT PORTRAIT;
*SRE
(Service Request Enable) sets and queries the bits in the Service Request Enable
Register (SRER). For a complete discussion of the use of these registers, see
page 3–1.
Status and Error
*CLS, DESE, *ESE, *ESR?, EVENT?, EVMSg?, FACtory, *PSC, *STB?
*SRE <NR1>
*SRE?
?
<Space>
<NR1>
*SRE
<NR1> is a value in the range from 0 to 255. The binary bits of the SRER are set
according to this value. Using an out-of-range value causes an execution error.
The power-on default for SRER is 0 if *PSC is 1. If *PSC is 0, the SRER
maintains its value through a power cycle.
*SRE 48
sets the bits in the SRER to 00110000 binary.
*SRE?
might return a value of 32, showing that the bits in the SRER have the binary
value 00100000.
*STB? (Query Only)
(Read Status Byte) query returns the contents of the Status Byte Register (SBR)
using the Master Summary Status (MSS) bit. For a complete discussion of the
use of these registers, see page 3–1.