Clear and sanitize procedures
To restore the factory
default settings
Restore the ins
trument to the factory default settings as follows:
1. Press the MODULE button until TG8000 : PRESET appe
ars.
2. Press the ENT E R button. You will see the words RECALL appear.
3. Press the left (◄)orright(►) arrow button until Factory Default appears.
4. Press the ENTER button.
5. Press the BACK button or the MODULE button to exit the Recall preset
menu.
To set the factory default
settings as the power on
state
1. Perform the previous procedure to restore the fa ctory preset/default state.
2. Press the MODULE button until TG8000 : PRESET appears.
3. Press the ENTER button. You will see TG8000 : PRESET : RECALL
displayed.
4. Press the up (▲)ordown(▼) arrow button to select TG8000 : PRESET
:SAVE
5. Press the left (◄)orright(►) arrow button until the display shows TG8000:
PRESET : SAVE Power On Default.
6. Press the ENTER button. The display will flas h as it saves the preset.
7. Press the BACK button to exit the preset menu.
To
clear the network
settings
1. Press the MODULE button until TG8000 : PRESET appears.
2. Press the up (▲)ordown(▼) arrow button to select TG8000 : UTILITY,
andthenpresstheENTER button.
3. Press the up (▲)ordown(▼) arrow button to select TG8000 : UTILITY
: NETWORK.
4. Press the left (◄)orright(►) arrow button to select Setup,andthenpress
the ENTER button.
5. Press the up (▲)ordown(▼) arrow button to select TG8000 : UTIL : NET
SETUP : DHCP.
6. Press the left (◄)orright(►) arrow button to select Disable.
NOTE. The DHCP feature must be disabled before you can clear the network
settings stored in the instrument memory.
7. Press the up (▲)ordown(▼) arrow button to select TG8000 : UTIL : NET
SETUP : IP ADDRESS, and then press the ENTER button.
TG8000 Declassification and Security Instructions 7