Clear and sanitize procedures
Table 1: Volatile memory devices for the TG8000 mainframe (cont.)
Type and
minimum
size Function
Type
of
user
data
Backed
up by
battery
Method
of
modication
Data
input
method Location
User
accessible To clear
To
sanitize
5M240ZTQFP100
CPLD
volatile
keypad
read-back
and LED
drive control
registers
Settings
No Indirectly UI or
remote
Front panel /
U1
Yes
Set power
on preset to
factory, cycle
power.
Repeat
clear
process
three
times.
Real Time
Clock +
128x8 ser
ial
NV memory
The RTC
function i
s not
accessible
but the clock
runs and
will show
the elapsed
time sin
ce the
board was
built. The
memory
is
used to store
user network
param
eters.
Data
and
settings
Yes Direct
and
indirect
UI or
remote
Main / U16
Yes
Perform
the clear
network
settings
procedur
e.
There is no
way to clear
the real
time
clock.
Remove
part or
battery.
2 TG8000 Declassication and Security Instructions

Other manuals for Tektronix TG8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TG8000 and is the answer not in the manual?

Tektronix TG8000 Specifications

General IconGeneral
BrandTektronix
ModelTG8000
CategoryPortable Generator
LanguageEnglish

Related product manuals