EasyManua.ls Logo

TEL AFA 1000 - Airflow Status: High, Sash High

TEL AFA 1000
22 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
High airflow
If configured :-
High Air toggles on / off with display
Audible alarm sounds – can be muted via Enter pushbutton)
Sash High
When the input configured as Sash High is activated
Amber LED on
Sash High – toggles on / off with velocity display
Audible alarm sounds
Audible can be muted via Enter pushbutton -- this silences the alarm and
initiates a repeat timer ( if configured ). After the delay time the alarm re-
sounds ( and can be re-muted). During this time the Amber LED flashes on /
off.
Sash High relay operates ( if configured )
Reset when Sash lowered to safe position and input de-activated.