Blood Pressure Monitor for
Medical Professionals from TERUMO
®
Getting Started
1. Check the unit
!
Battery is in properly
!
Tubing is inserted all the way in
*
!
Ination bulb is screwed in properly
2. Select the right cuff size
!
Make sure to select the right size
cuff checking the reference lines
3. Wrap the cuff
!
The “O” mark needs to be placed
right above the artery
Normal & Slow Modes
Auscultation Mode
4. Turn the power on
(Defaults to Normal
mode. Choose “Slow”
for hypotensive patients.)
4. Turn the power on
and choose Aus”
6. Stop pressurization
(measurements
start automatically:
SYS, DIA, & pulse
will display when
complete)
7. Press Pow er button
to turn off monitor
(automatically shuts off
after 3 idle minutes)
5. Pressurize
30-40 mmHg
above expected
SYS pressure
using bulb
5. Pressurize
30-40 mmHg
above expected
SYS pressure
using bulb
6. Stop pressurization
(No automatic read-out
in Auscultation)
7. Press Pow er button
to turn off monitor
(automatically shuts off
after 3 idle minutes)
*
Site window
Question and Answer IconNeed help?

Do you have a question about the Terumo Elemano and is the answer not in the manual?

Terumo Elemano Specifications

General IconGeneral
Product NameTerumo Elemano
CategoryBlood Pressure Monitor
DisplayDigital LCD
Measurement MethodOscillometric
Measurement Range (Pressure)0-299 mmHg
Accuracy (Pressure)±3 mmHg
Accuracy (Pulse)±5%
Cuff Size22-32 cm
Measurement Range (Pulse)40-180 bpm
Power SourceBattery
Number of Batteries4
Memory60 measurements

Summary

Getting Started

Check the Unit

Verify battery, tubing, and inflation bulb are properly connected and inserted.

Select Right Cuff Size

Ensure the correct cuff size is selected by checking reference lines for proper fit.

Wrap the Cuff

Apply the cuff correctly, ensuring the 'O' mark is positioned directly above the artery.

Operating Modes

Normal & Slow Modes Operation

Turn on the monitor, pressurize the cuff 30-40 mmHg above expected systolic, and stop pressurization.

Auscultation Mode Operation

Turn on the monitor, select 'Aus', pressurize the cuff, and stop pressurization without automatic read-out.

Troubleshooting Common Issues

Low Indication

Indicates insufficient pressurization; recommended to pressurize 30-40 mmHg higher than expected systolic.

Difficult Pressurization

May be due to a clogged air filter or deteriorated bulb; cleaning or replacement may be needed.

Flashing Panel Warnings

Panel flashes to warn of abnormal conditions like irregular beats; use Auscultation mode to verify.

Error Code Troubleshooting

Addresses errors like Err 1, 2, 3, 5 (connection issues) and Err 4 (abnormal readings) with specific measures.

Cuff Size Guide

Arm Circumference Sizing

Provides a chart correlating cuff sizes (SS, S, M, L) with arm circumference in centimeters and inches.