7 Technical data
21
• Adjustment conditions for flow probes:
Adjustment in free jet Ø 350 mm, reference pressure 1013 hPa,
based on testo reference Laser Doppler Anemometer (LDA).
• Note for flow probes:
Low flow velocities may result in greater measurement
uncertainties during humidity and temperature measurement!
• Note for humidity probes:
Please do not use the humidity probes in condensing atmospheres.
For continuous use in high-humidity ranges
> 80% RH at ≤ 30°C for > 12 h
> 60% RH at > 30°C for > 12 h
please get in touch with Testo Service or contact us via the Testo
website.
Bluetooth
®
handle for connecting testo 440 probe heads (0554 1111)
Feature Value
Dimensions 170 mm x 50 mm x 40 mm
Warranty Duration: 2 years
Hot wire probe with Bluetooth
®
, including temperature and humidity sensor
(0635 1571)
Feature Value
Measuring range 0 to +50 m/s
-20 to +70°C
5 to 95% RH
Accuracy
(at 22°C, ±1 digit)
±(0.03 m/s + 4% of m.v.) (0 to 20 m/s)
±(0.5 m/s + 5% of m.v.) (20.01 to 30 m/s)
±0.8°C (-20 to 0°C)
±0.5°C (0 to +70°C)
w ww. . com