5-2 3177885 Thelco Oven Thermo Electron Corporation
Section 5
Installation
Caution Be sure that the power supply is of the same voltage as specified
on the nameplate.
V
Caution Be sure the wall receptacle is readily identifiable and easily reached
to disconnect the unit from the power source.
V
Model 6557 115-volt units (Cat. No. 3166789) are equipped with non-
standard plugs (one horizontal and one vertical blade) and will require
installation of a special matching receptacle. To be in compliance with the
National Electrical Code, non-standard plugs and receptacles are required
to accommodate higher current draw.
The aforementioned models are rated at 15.6 amps.