EasyManua.ls Logo

Thermo Forma 3940 - 1900140 IR CO Option

Default Icon
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Models 3940 and 3949 ________________________________________________________________Installation and Set-Up
2.12 1900140 IR CO2 Option
This section applies to units with the 1900140 IR CO
2
option only.
a. Connecting the CO
2 Source (Refer to Figure 2-3)
For the most economical use, the liquid CO
2 supply tanks
should be without siphon
tubes, so that only CO2 gas
enters the incubator injection
system. Two tanks may be
joined together with a mani-
fold to ensure a continuous
CO
2 supply.
Install a two-stage pres-
sure regulator, with indicating
gauges, at the supply cylinder
outlet. The high-pressure
gauge should have an indicat-
ing range of 0 to 2000 psig to
monitor tank pressure. The
low-pressure gauge should
have an indicating range of 0
to 30 psig to monitor input
pressure to the incubator injection system. A suitable two-stage
pressure regulator is available.
The CO
2 source must be regulated at a pressure level
between 5 and 10 psig. Higher pressure levels may damage the
CO2 control system. The user should determine the most eco-
nomical pressure level, between 5 and 10 psig appropriate for
the desired CO
2 percentage in the chamber. Use only sufficient
pressure to maintain recovery time after door openings.
To connect the CO
2 supply:
· Insert the copper tubing provided with the unit as far as it
will go into the nut of the CO2 connection.
· Turn the nut until it is finger tight.
· For reference, scribe the nut at the 6:00 position.
· While holding the fitting body steady with a wrench, tighten
the nut 1-1/4 turns until the mark is at the 9:00 position.
· Securely attach the CO
2 line to the open end of the copper
tubing.
· Check the tubing connection for leaks.
b. Setting the CO
2
Content
The Watlow 93 CO
2 controller's upper display shows
the actual CO2 content inside the chamber. The lower display
shows the CO2 setpoint.
Before setting the CO
2 Content, allow the chamber tempera-
ture and humidity to stabilize. Do not open door during the sta-
bilization period.
To set the CO
2 content, press the up or down triangles
on the Watlow 93 Controller.
c. CO
2 Controller Calibration
If it should become necessary to calibrate the CO
2 con-
troller, perform the following procedure. From the standard
operating display (setpoint in bottom display, actual CO
2 reading
in the upper display):
Enter the Setup Menu by pressing the t up-arrow and
6down-arrow keys
simultaneously on the
Watlow 93 controller for
three seconds. The
lower display shows the
Lock parameter, and the
upper display shows its
current level. Press
the6 down-arrow until
0 is displayed. Then,
use the Advance key to
move through the menus
until the "CAL" prompt appears.
Once at the "CAL" prompt, press thet up-arrow or 6
down-arrow to add or subtract the percentage of CO
2 to make
the display accurate. Press the Advance key until LOC appears.
Press the t up-arrow key to place "2" in the upper display.
Continue to press the Advance key until the setpoint display
appears.
d. CO
2
Control and Indicators
Sample Port (Refer to Figure 2-9)
The sample port is used for checking CO
2 percentage in the
incubator chamber by an independent test instrument (such as
with a Fyrite, or similar CO2 test instrument).
To prevent CO2 loss, the sample port must be capped when it is
not in use.
CO2 Alarm (Refer to Figure 2-9)
The CO2 alarm is factory set to activate when the
chamber CO
2 content
deviates ± 5%. When a
CO2 alarm occurs, the
CO
2 Alarm indicator on
the control panel lights
and the audible alarm
sounds.
The CO2
alarm high and low set-
points are established
through the Watlow 93
CO2 controller (ALO,
and AHI). Refer to the
configuration record
included at the end of
Section 2.
2 - 5
Figure 2-7
Figure 2-8
Figure 2-9