What to do if Thermo Scientific Security Sensors detect leakage?
- PpramosAug 20, 2025
If the leak sensor detects leakage, you should find and eliminate the source of the leakage.
What to do if Thermo Scientific Security Sensors detect leakage?
If the leak sensor detects leakage, you should find and eliminate the source of the leakage.
What to do if Thermo Scientific Security Sensors show low gas pressure?
Low gas pressure can be caused by an incorrect gas pressure setting, a gas leak, or an incorrect gas supply pressure. Ensure that the gas supply pressure is within the allowed range.
What to do if Thermo Scientific Security Sensors show high flow ratio?
If you observe a high flow ratio, ensure that gas is still supplied to the detector. Also, check that the gas supply pressure is within the allowed range and observe the gas specifications for the detector.
What to do if Thermo Scientific Corona Veo show liquid sensor fault?
If a liquid sensor fault is detected, it means excess liquid is present inside the detector. Make sure that the gas flow and drain pump are turned on and check that liquid flows out of the waste line.
What to do if Thermo Scientific Security Sensors show high evaporation temperature?
If there is a high evaporation temperature, it means the internal temperature of the evaporation tube is too high. Set a lower evaporation temperature.
What to do if Thermo Scientific Security Sensors show low flow ratio?
If a low flow ratio is detected (the gas flow ratio is below the specified range for more than 60 seconds), make sure that gas is still supplied to the detector and check that the gas supply pressure is within the allowed range.
What to do if Thermo Scientific Security Sensors report 'Charger voltage failed'?
If the charger voltage fails because the voltage of the charger approaches the maximum level, turn the detector off and on.
What to do if Thermo Scientific Security Sensors show 'The internal temperature of the engine is too high'?
If the internal temperature of the engine is too high, turn the detector off and on.
What to do if Thermo Scientific Corona Veo show 'The internal temperature of the engine heater is too low'?
If the internal temperature of the engine heater is too low, turn the detector off and on.
What to do if Thermo Scientific Corona Veo Security Sensors show 'Evaporation heater fault'?