Guide to Replacing
the Modular Cable
of the Driveline
for Clinicians
For additional information
and guidelines, please refer to the
HeartMate 3 Instructions for Use.
Thoratec Corporation 6035 Stoneridge Drive, Pleasanton,
CA 94588 Tel: (925) 847-8600, Fax: (925) 847-8574,
HeartLine: (800) 456-1477, www.thoratec.com
Thoratec Corporation continually strives to provide the highest quality products for mechanical circulatory support.
Specications may change without notice. Therefore, please refer to the HeartMate 3 Left Ventricular Assist System IFU
shipped with the HeartMate 3 for the most current information regarding indications, contraindications, warnings, and
cautions. Thoratec, the Thoratec logo, and HeartMate are registered trademarks and HeartMate 3 and HeartLine are
trademarks of Thoratec Corporation.
©2017 Thoratec Corporation. 10006921.A 08/2017
Option 2: Exchange Modular Cable Only
IMPORTANT! Ability to successfully change Modular Cable may be affected by
several factors, such as native cardiac output, cognitive function, etc., which may change over the course
of LVAD support, and therefore should be periodically reassessed.
a. Open the Safety Lock on the System Controller.
b. Loosen the Safety Nut on the Modular In-Line Connector until the clicking sound has stopped and it
spins freely.
a. Press the red button and remove the Controller Driveline Connector of the current Modular Cable.
The pump will stop during this time.
b. Align the WHITE arrow/ alignment mark on the Driveline with the WHITE arrow on the Controller.
c. Insert the Controller Driveline Connector of the replacement Modular Cable into the System Controller.
d. Close the Safety Lock on the replacement System Controller. The Safety Lock cannot move to the
locked position unless the driveline is fully and properly inserted.
a. Disconnect the Modular In-Line connector on the current Modular Cable by pulling the connector apart.
b. Insert the replacement Modular Cable into the Modular In-Line connector. Once connected, the pump
will restart within 10 seconds. Rotate the Locking Nut closed until the clicking stops and the yellow line
is hidden by the Locking Nut.