E6
E7
E9
Alarm condition: The door handle was not turned to the end position, and the door handle
turn back when pressure rising.
Alarm condition: Overtimes
Alarm condition: Holding temperature is failed
Solution: Turn the door handle to the end and run a cycle again. If it still appears E6 again, please
check if the door goes down, if yes, the door has to be readjusted. To make the door be horizontal.
Solutions: the same as E4
Case 1: Adjust the parameter refer the instructions manual if in a high altitude place.
Case 2: Check the heater of steam generator. One of the heaters does not work.
Case 3: Check if there is a big leakage on the steam generator. Or in the other place
Case 4: Check the water release valve V4 and air release valve V1, if they can be closed completely.
You may observe the tube before the valve when the program is running. If the valve have
not been closed completely you will find the steam or water flow to condenser continuously.
EF
Alarm condition:
System have not initialized
Usually it is caused by the interference from the power net. The data saved in the chip is lost.
We need to reset the data. And after reset the data. We recommend using the manostat to avoid
interference.
7
Correct the parameter of sensor for 18/23-A
1. Power off
2. Open the cover of machine
3. Replace the 3 temperature sensors by the standard sensor.
4. Power on. And open the door.
5. Holding the SET button until you enter the menu. Release the button.
6. Select PT by press the SELECT button.
7. Press SET button. You will heard one long sound and four short sound.
That means the operation is success.
8.Press START button. Power off and replace the sensors.