Do you have a question about the Titan Logix TD 100 and is the answer not in the manual?
Brand | Titan Logix |
---|---|
Model | TD 100 |
Category | Transmitter |
Language | English |
Establish connection to the TD100 transmitter via direct or Finch II methods.
Launch and select the TD100 utility within the SensorLink software.
Configure probe type, detection mode, depth, and volume units.
Manually enter depth and volume values into the strapping table.
Enter tank thickness and probe offset measurements for calibration.
Set spill alarm levels, reset modes, and HH alarm thresholds.
Define display resolution and decimal places for volume measurement.
Configure sump depth and 4-20mA points if applicable to the installation.
Check the entered strapping table data for errors before saving.
Save the configured strapping table to a file for future use.
Upload the strapping table and settings to the TD100 transmitter.
Load a previously saved .stb strapping table file into SensorLink.
Import strapping table data from a .csv file format into SensorLink.
Transfer configuration from a TD80 transmitter to a TD100 transmitter.
Adjust rotary switches on the Finch II enclosure to match display settings.
Configure the fill alarm level using the Finch II display interface.
Configure the fall alarm level using the Finch II display interface.
Power on the TD100 system and verify initial display and relay outputs.
Acknowledge and clear active alarms displayed on the Finch II.