172-65592MA-07 (TM5N/TM5N-EX TrapMan Type P) 12 May 2022
1. Read Carefully Before Use
1.1 Periodic Required Maintenance
In order for TM to perform highly accurate diagnosis, a periodic function check is necessary.
TM will use the following graded alert system to remind users of the upcoming scheduled
maintenance. When one of the following alerts is displayed on screen, as soon as possible,
bring the unit to the nearest TLV representative, or contact TLV for further information.
1. The TM begins counting when you turn on the power for the
first time (or after completion of maintenance), and when the
2 years until the required maintenance plus a 3 month grace
period have passed, the unit will be locked.
Starting 3 months before the next scheduled maintenance, a message showing the
number of days left until required maintenance, will be displayed every time the unit is
turned on.
Press
to continue with normal operation.
2. If the scheduled date for maintenance has passed and
maintenance still has not been carried out, for a further 3
months a message will be displayed showing the number of
days that maintenance is overdue and the number of days
until the TM unit becomes locked.
Press
to continue with normal operation.
3. If maintenance is not carried out before the lock date (2 years
and 3 months from first power-on), a message will be
displayed stating that the unit is currently locked.
When the unit is locked you will only be able to:
Turn the power ON/OFF,
Turn the light on or off,
Transfer data to and from the TrapManager software.
All other functions will be unavailable.
DAYS UNTIL REQ.
MAINTENANCE: XXX
MAINT OVERDU: YYY
DAYS TO LOCK: ZZZ
MAINT OVERDU: YYY
UNIT LOCKED