- 10 -
TOTOKU ELECTRIC CO., LTD.
MQSA Quality Control Manual
5 IMAGE QUALITY CHECK - SMPTE PATTERN
5.1 Objective
To calibrate the display monitors as accurate as possible for the reading of
mammographic images on Monochrome Monitors for Mammography
5.2 Frequency
Daily
5.3 Required Equipment
Monochrome Monitors for Mammography, display monitor
Serial cable for communicating between PC and display monitor (for ME511L)
USB cable for communicating between PC and display monitor (for other models)
Calibration software (Medivisor Nx)
5.4 Procedure
1. Select “TOTOKU Medivisor Nx”>”Medivisor Nx” from the Start menu to start this
software application.
2. Select the “Test Pattern” Tab. Click the display monitor icon to choose the
monitor to test and click the “SMPTE” pattern.