EN-14 FBx Series Bench Scale
If calibration fails, please press key to return, then to check if
parameter setting is correct.
Step 8:Turn off the power, then turn on the power again, place a known
weight on the platform to verify if it is accurate.
Step 9: If the weight displayed does not correspond with the value of the
calibration weight, please repeat above steps.
Step 10: After calibration is completed, open the CAL jumper of J13,
re-install the front cover and make sure the liquid-tight connector is tight.
Menu flow chart is shown below