TRITEMP™ Non-contact thermometer
EN
15
XI. Adjustment of MODE
settings
The device has 4 MODE options as
follows:
1. Unit of temperature °C or °F
2. Audio enabled ON / OFF
3. READ Mode BODY / SURFACE TEMP
4. OFFSET > +3 to -3 °C / +5.4 to
-5.4°F.
• The device is supplied with factory
settings and the device screen will
display °C , Audio ON, BODY and
0.0 offset.
• Alteration of the factory settings
must only be performed by a
qualiedBiomedicaltechnician
which are aligned with the
requirements of the clinical team.
• To adjust any MODE settings press
and hold the POWER button until
you hear an audible bleep. (See
Figure 1. Button Functions, Pg 10).
Release POWER button and °C will
startashingonthescreen.Theuser
can the move through each MODE
option (from 1 to 4 as listed above)
by pressing the POWER button each
time.
SETTING THE UNIT TEMPERATURE
(°C OR °F)
• To change the unit of measurement
from °C to °F
o Press and hold the POWER button
until you hear an audible bleep.
o Release POWER button and °C will
startashingonthescreen.
o Press the READ button to change
to°F.The°Fwillstarttoashon
the display.
o Press the POWER button 4 times
to move through the MODE
options and switch off the device
• To change the unit of measurement
from °F to °C
o Press and hold the POWER button
until you hear an audible bleep.
o Release POWER button and °F will
startashingonthescreen.
o Press the recessed adjust button
with the tip of a paperclip to revert
back to °C.
o Press the POWER button 4 times
to move through the MODE
options and switch off the device.
The device will display °C when
turned ON.