Chapter 2
4
Setting up the Model 8026 Particle Generator
The particle generator supplements the naturally occurring particles in the
room environment. Follow the steps below to set up the particle generator:
1. Fill the reservoir jar with clean tap water to the fill line. It is helpful use
hot water to help the salt tablet dissolve more quickly.
Figure 2-1
2. Drop one salt tablet into the reservoir. Each tablet contains 1.0 grams of
salt. Break the tablet before dropping it into the water to dissolve it
more quickly.
3. Place the temporary cover tightly on the reservoir and shake it gently to
dissolve the salt tablet.
4. Once the solution is mixed, remove the temporary cover and screw the
reservoir onto the particle generator. See Figure 2-1.
5. Place the particle generator on the floor in a corner of the testing room.
Make sure the ventilation louvers on the back and bottom of the
generator are not obstructed, as this could cause the instrument to
overheat.
Caution
!
Do not place the Particle Generator on the table or
benchtop. Vibrations from the internal compressor may
cause the generator to move across the table. If left
unattended, the generator could fall on the floor resulting
in damage to the instrument.
Fill line