EasyManua.ls Logo

TSI Instruments AEROTRAK+ 7201 - 4-20 Ma Analog Output; Wireless Connector; Tubing Connections; Inlet

TSI Instruments AEROTRAK+ 7201
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24 AeroTrak
®
+ Remote Particle Counter Operation Manual
4-20 mA Analog Output / Wireless Connector
This 6-pin connector is used for either a 3-channel 420 mA output or to
connect to a TSI wireless node, depending on the model used.
Two (2) of the analog
outputs are configurable, via
the Configuration Utility, to
linear/log scale and linked to
a particle channel.
Terminal
Channel
Direction
1
A
Output
2
A
Return
3
B
Output
4
B
Return
5
Status
Output
6
Status
Return
The last analog output is
dedicated to status
information. The output will
go to a nominal current value
that corresponds to the
condition listed in table.
Current
Flow
Status
Laser
Status
8 mA
OK
OK
12 mA
OK
Alarm
16 mA
Alarm
OK
20 mA
Alarm
Alarm
Tubing Co n necti o n s
Inlet
The direct-mount sampling inlet at the top of the device can be
configured to sample standalone, connected to a sample tube, or
connected to an isokinetic sampling inlet. See the table of optional
accessories in Chapter 1 for more information.
0.1 CFM Inlet 1.0 CFM Inlet