Chapter 4
Maintenance
The Alnor Model AXD620/AIRFLOW PVM620 requires very little
maintenance to keep it performing well.
Recalibration
To maintain a high degree of accuracy in your measurements, we
recommend that you return your Alnor Model AXD620/A
IRFLOW PVM620
to TSI for annual recalibration. Please contact one of TSI’s offices or your
local distributor to make service arrangements and to receive a Return
Material Authorization (RMA) number. To fill out an online RMA form,
visit TSI’s website at
http://service.tsi.com.
U.S. & International UK
TSI Incorporated
TSI Instruments Ltd.
Alnor Products Lancaster Road
500 Cardigan Road Cressex Business Park
Shoreview MN 55126-3996 High Wycombe
Tel: (800) 874-2811 Buckinghamshire
(651) 490-2811 HP12 3QP England
Fax: (651) 490-3824 Tel: +44 (0) 149 4 459200
Fax: +44 (0) 149 4 459700
The Alnor Model AXD620/A
IRFLOW PVM620 can also be recalibrated in
the field using the CALIBRATION menu. These field adjustments are
intended to make minor changes in calibration to match a user's calibration
standards. The field adjustment is NOT intended as a complete calibration
capability. For complete, multiple-point calibration and certification, the
instrument must be returned to the factory.
Cases
If the instrument case or storage case needs cleaning, wipe it off with a soft
cloth and isopropyl alcohol or a mild detergent. Never immerse the Alnor
Model AXD620/A
IRFLOW PVM620. If the enclosure of the Alnor Model
AXD620/A
IRFLOW PVM620 or the AC adapter becomes broken, it must be
replaced immediately to prevent access to hazardous voltage.
Storage
Remove the batteries when storing the unit for more than one month to
prevent damage due to battery leakage.