10 Ohm mode setting calibration
Select:
Enable = out, Mode = 10 Ω.
Meters = Level A & Level B, check Level A setting is 5.00 Ω.
Set: Supply output to 15 Volts (current required: 3 Amps).
Select: Enable = in. Meters = Watts & Ohms.
Check: on scope that the system is stable.
Adjust: VR16 for Ohms (RH) meter to read 5.000
31