-29-
Problem Cause Measures
Reference
The pump is too small for the
volume of the vacuum chamber.
Select another pump. 5.2
The pressure measurement
method is not correct.
Measure the pressure correctly. 5.1
The vacuum gauge is not
suitable.
Measure with a calibrated
vacuum gauge suitable for the
pressure range.
5.1
The pipe connected to the inlet
port is small, or the piping
distance is long.
Use pipes having a diameter
larger than the inlet port
diameter, or reduce the distance
from the vacuum.
5.1
The wire mesh at the inlet port is
clogged.
Remove the piping from the
upper section of the inlet port,
and clean the wire mesh.
6.2
The specified amount of oil has
not been added.
Add the specified amount of
oil.
6.3
The oil has deteriorated. Replace the oil. 6.3
Leakage occurs from the pipe
connected to the pump.
Locate the leakage with a
leakage detector and stop the
leakage.
--
Our specified oil is not being
used.
Overhaul the pump and replace
with oil specified by us.
6.3
The pressure does
not decrease.
Oil does not circulate, or the
oil hole of the cover is clogged.
Overhaul and clean the oil hole. 6.2
Problem with power supply
voltage.
Set the power supply voltage to
within ±10% of the rated
voltage.
3.5
The motor malfunctions. Replace the motor. 6.2
Foreign matter has entered the
pump.
Eliminate the foreign matter
and overhaul the pump.
6.2
The specified amount of oil has
not been added.
Add the specified amount of
oil.
6.3
Abnormal sound is
generated.
The coupling malfunctions. Overhaul (replace the
coupling).
6.2