-30-
Problem Cause Measures
Reference
Abnormal sound is
generated.
Oil does not circulate, or the oil
hole of the cover is clogged.
Overhaul and clean the oil hole. 6.2
Components inside the pump
have burnt out.
Overhaul (replace the damaged
components).
6.2
Pump surfaces are
extremely hot
Continuous operation at high
evacuation pressure.
When it runs by high inhalation
pressure consecutively, there is
no problem in particular in
pump surface temperature to
less than 40K.
4.1
The specified amount of oil has
not been added. (If the oil
amount is not sufficient, the
cooling effect
of the pump will be reduced.)
Add the specified amount of
oil.
6.3
The temperature of the evacuated
gas is high.
Mount cooling equipment such
as a gas cooler at the inlet side.
--
Oil does not circulate, or the oil
hole of the cover is clogged.
Overhaul and clean the oil hole. 6.2
A lot of oil splashes
out from the outlet
port.
The pump is been filled in excess
of the specified amount.
Discharge the oil until it reduces
to the specified amount.
6.3
Continuous operation is
performed at a high evacuation
pressure.
Install an oil mist trap at the
outlet side.
4.7
4.8
The oil leaks
outside the pump.
Deterioration of the O-ring and
the oil seal of the case and cover.
Check and replace the O-ring
and oil seal.
6.2