D/N.MFST10CE,2022DR08 ULVAC CRYOGENICS INCORPORATED
SUPER TRAP TEMPERATURE
CONTROLLER
STC-1C series
Instruction Manual
STC-1C
STC-1C(INV)
Export Control Policy
We recommend that ALL customers be sure to follow all
rules and regulations such as Foreign Exchange and
Foreign Trade Law when exporting or reexporting our
products.

Questions and Answers

  • A
    anthony88Aug 8, 2025
    What to do if the Ulvac STC-1C refrigerator unit does not start running?
    • C
      Crystal SuttonAug 8, 2025
      If the refrigerator unit of your Ulvac Temperature Controller doesn't start, possible causes include disconnected input power, refrigerator power, or inverter control cables; contact failure in these cables; a compressor failure; a refrigerator failure; or an internal circuit fault. Ensure all cables are correctly connected. If the issue persists, it may indicate a failure of the compressor or refrigerator. In that case, check their instruction manuals.
  • J
    Jessica Nash DDSAug 15, 2025
    Why can't I set the parameter on my Ulvac STC-1C Temperature Controller?
    • K
      Kimberly HubbardAug 15, 2025
      If you can't set a parameter on your Ulvac Temperature Controller, it's likely because the function to inhibit parameter setting is activated. To release this function, press the [UP] and [DN] keys simultaneously for 5 seconds.
  • K
    Kevin RobinsonAug 23, 2025
    What to do if the Ulvac STC-1C display shows [400K] and does not move?
    • L
      lorimitchellAug 23, 2025
      If the display of your Ulvac Temperature Controller shows [400K] and doesn't change, it could be due to the K thermocouple or its cable not being connected, or the K thermocouple being disconnected. Ensure the K thermocouple and its cable are correctly connected. If the problem continues, replace the K thermocouple.
  • J
    jennifer68Aug 29, 2025
    Why has the indicated temperature on my Ulvac Temperature Controller reached room temperature or above (around 340K)?
    • K
      Karl AcostaAug 30, 2025
      If the indicated temperature on your Ulvac Temperature Controller is at or above room temperature (around 340K), it could be due to incorrectly wired [?] and [?] of the K thermocouple, or interference from a heater or another heat source. Make sure the thermocouple wires are correctly oriented. Immediately cut off any nearby heater sources, as the refrigerator's maximum allowable temperature is 340K.

Summary

Introduction, Safety, and Policies

Export Control Policy

Policy regarding export and reexport of products.

Product Introduction and Personnel Requirements

Overview of the manual and safety training requirements for personnel.

Safety Considerations

Explains general safety symbols, warnings, and precautions for safe operation.

Product Overview and Specifications

Features, Types, and Specifications

Covers product features, models, and detailed technical specifications.

Components and Connectors

Describes unit components, controls, and connector pinouts for STC-1C and STC-1C(INV).

Accessories and Optional Cables

Lists available accessories and optional cables for the STC-1C series.

Installation and Wiring

Installation Guidelines

Instructions for selecting installation site and physically mounting the unit.

Wiring Procedures and Precautions

Guidelines and precautions for proper wiring, including diagrams and conduit requirements.

Operation, Maintenance, and Troubleshooting

Operation and Parameter Setting

Steps for operating the unit, setting parameters, and initial configuration.

Troubleshooting Common Issues

Common problems, possible causes, and actions for resolving issues.

Service Network and Revisions

Information on service network and manual revision history.