Caution: DISCONNECT THE POWER CORD BEFORE REPLACING FUSE.
Follow the procedures below to replace the fuse:
a) Unplug the power cord. Turn the on/off switch to “off”
b) Remove the power cord from the power inlet on the back of the centrifuge.
c) Locate the fuse holder. The fuse holder is a part of the power inlet.
On/off switch Inlet
Fuse Holder
d) Use a Philip type screwdriver to take the snap-in type holder cover off.
Remove the blown fuse and replace it with the same type and rating fuse
which is:
250V 4A
e) Put the snap-in fuse holder back.
5. Intermittent motor operation or no operation:
Contact your representative for possible motor replacement.
Specifications:
Speed: 1,000 - 3,400 RPM
Speed accuracy: +/- 25 RPM
Digital Timer: 1-60 Minutes
Rotor: 4 Place Horizontal Spin
Capacity: Four x 50ml
Dimension (L x W x H): 13 ¾ ” x 15” x 11”